Pharmaceutical Isolators

Aseptic Isolator, Aseptik İzolator

Isolators work under the principle of particle filtration, air recirculation and keeping work chamber at positive or negative pressure. When isolator is at positive pressure, 90% (±%5) of total internal air is recirculated and passes through main HEPA/ULPA filter located at work zone top. The rest of the air is exhausted after passing through the HEPA/ULPA filter located at the top of the cabinet. Negative pressure isolators exhaust total cabinet air completely to the outer atmosphere without recirculation after its passing through the main filters. Pre/HEPA filtered room air is taken in by the top of the cabinet and pass-box by providing an ISO-5 class work zone. In addition to the primary air barrier accomplished by within the working zone, double-wall negative plenum design of the Metisafe cabinet prevents particle from escaping to surroundings and zero leakage is ensured.

Main Body & Design
  • Antibacterial epoxy powder paint coated aluminum
  • 70 angle sloped front panel for work comfort and increased working chamber vision
  • Isolated automation systems from work chambers
  • Interlocked ISO-5 class dynamic pass-box with sliding load tray
  • UV resistant safe front glass & radiused cornered pass-boox door glasses
  • Leakproof 0,6 > mm chemical resistant gloves
Decontamination System
  • Full automatic decontamination process working parallel with work chamber temparature and humidity level
  • Integrated H2O2 pulverization
  • Sensitive H2O2 measurement
  • Timer controlled efficient UV disinfection
Microprocessor Control Unit 
  • Air velocity
  • Password protection
  • Normal/Stand-by work mode option
  • HEPA Filter Life, filter integrity status
  • Alarm activation log data
  • Work-chamber temperature and humidity rates
  • Illumination level adjustment
  • Full automatic decontamination and process follow-up screens
  • UV and FL Lamp work times
  • Online particle count data (optional)
Audio-Visual Alarm
  • Unproper pressure warning
  • System ready/not-ready warning
  • Service need warning
  • Alarm cancel button
  • Filter/Lamp replacement warning
Quality and Certificates
  • CE
  • Management certificate of ISO9001-2015
  • Biological Safety Cabinets EN 12469
  • ISO 14644 Clean Room class compatibility
  • EN61010 Electrical and Electronic safety compatibility
  • Validation guarantee from the accredited test company
Internal Dimensions (WxLxH) mm1250x620x695
External Dimensions (WxLxH) mm1950x902x1705
Support Stand (WxLxH) mm1371x805x745  
Work Table Height840 mm
Work Chamber Pressure - Positive + 20 Pa
Filters (EN 1822)Main FilterH14 HEPA, 0,3 µm %99.995 < 
Exhaust FilterH14 HEPA, 0,3 µm %99.995 < 
Pre-FilterEU4
Work Chamber Cleanroom ClassEN ISO 14644-3< ISO 5
FED209E< Class 100
Work BenchStandard304 Stainless Steel
Optional316L Stainless Steel
Noise LevelNormal Mode 56 dB(A)
Stand-by Mode 50 dB(A)
Front Glass≤ 10 mm Laminated
Front Glass Gloves Port Quantity2
Work Chamber Illumination (Lux)850 – 1250 Lux
Energy consumption
230 VAC - 50 Hz
Fan Filter Unit170 W
Exhaust Motor90 W
Illumination36 W
UV Lamp40 W
Decontamination Unit2800 W
Electrical Socket Power / Current2x1150 W / 2x5 A
Total Power / Current5436 W/23.6 A
Energy Consumption in Normal Operating Mode (Lighting and Fan Motor)296 W
Cabinet Weight (kg)340
Packaged Weight (kg)405
Pack/Palette Dimensions                           (Included Pass-Box and Support Stand ) mm2100x1150x2000
Internal Dimensions (WxLxH) mm1555x620x695
External Dimensions (WxLxH) mm2255x902x1705
Support Stand (WxLxH) mm1676 x805x745
Work Table Height840 mm
Work Chamber Pressure - Positive+ 20 Pa
Filters (EN 1822)Pre-FilterEU4
Exhaust FilterH14 HEPA, 0,3 μm particle %99.995< filtration efficiency
Main FilterH14 HEPA, 0,3 μm particle %99.995< filtration efficiency
Work Chamber Cleanroom ClassEN ISO 14644-3< ISO 5
US FED 209E< Class 100
Work BenchStandard304 Stainless Steel
Optional316L Stainless Steel
Noise LevelNormal Mode 56 dB(A)
Stand-by Mode 50 dB(A)
Front Glass≤ 10 mm Laminated
Front Glass Glove Port Quantity4
Illumination850 – 1250 Lux
Energy consumption
230 VAC - 50 Hz
Fan Filter Unit200 W
Exhaust Motor90 W
Illumination36 W
UV Lamp40 W
Decontamination Unit2800 W
Electrical Socket Power / Current2x1150 W / 2x5 A
Total Power / Current5466 W/23.7 A
Energy Consumption in Normal Operating Mode (Lighting and Fan Motor)326 W
Cabinet Weight (kg)400 
Packaged Weight (kg)475 
Pack/Palette Dimensions                                  (Included Pass-Box and Support Stand )2405x1150x2000
Metisafe-MSP-II-Teknik