Isolators work under the principle of particle filtration, air recirculation and keeping work chamber at positive or negative pressure. When isolator is at positive pressure, 90% (±%5) of total internal air is recirculated and passes through main HEPA/ULPA filter located at work zone top. The rest of the air is exhausted after passing through the HEPA/ULPA filter located at the top of the cabinet. Negative pressure isolators exhaust total cabinet air completely to the outer atmosphere without recirculation after its passing through the main filters. Pre/HEPA filtered room air is taken in by the top of the cabinet and pass-box by providing an ISO-5 class work zone. In addition to the primary air barrier accomplished by within the working zone, double-wall negative plenum design of the Metisafe cabinet prevents particle from escaping to surroundings and zero leakage is ensured.
Main Body & Design |
|
Decontamination System |
|
Microprocessor Control Unit |
|
Audio-Visual Alarm |
|
Quality and Certificates |
|
Internal Dimensions (WxLxH) mm | 1250x620x695 | |
External Dimensions (WxLxH) mm | 1950x902x1705 | |
Support Stand (WxLxH) mm | 1371x805x745 | |
Work Table Height | 840 mm | |
Work Chamber Pressure - Positive | + 20 Pa | |
Filters (EN 1822) | Main Filter | H14 HEPA, 0,3 µm %99.995 < |
Exhaust Filter | H14 HEPA, 0,3 µm %99.995 < | |
Pre-Filter | EU4 | |
Work Chamber Cleanroom Class | EN ISO 14644-3 | < ISO 5 |
FED209E | < Class 100 | |
Work Bench | Standard | 304 Stainless Steel |
Optional | 316L Stainless Steel | |
Noise Level | Normal Mode | ≤ 56 dB(A) |
Stand-by Mode | ≤ 50 dB(A) | |
Front Glass | ≤ 10 mm Laminated | |
Front Glass Gloves Port Quantity | 2 | |
Work Chamber Illumination (Lux) | 850 – 1250 Lux | |
Energy consumption 230 VAC - 50 Hz | Fan Filter Unit | 170 W |
Exhaust Motor | 90 W | |
Illumination | 36 W | |
UV Lamp | 40 W | |
Decontamination Unit | 2800 W | |
Electrical Socket Power / Current | 2x1150 W / 2x5 A | |
Total Power / Current | 5436 W/23.6 A | |
Energy Consumption in Normal Operating Mode (Lighting and Fan Motor) | 296 W | |
Cabinet Weight (kg) | 340 | |
Packaged Weight (kg) | 405 | |
Pack/Palette Dimensions (Included Pass-Box and Support Stand ) mm | 2100x1150x2000 |
Internal Dimensions (WxLxH) mm | 1555x620x695 | |
External Dimensions (WxLxH) mm | 2255x902x1705 | |
Support Stand (WxLxH) mm | 1676 x805x745 | |
Work Table Height | 840 mm | |
Work Chamber Pressure - Positive | + 20 Pa | |
Filters (EN 1822) | Pre-Filter | EU4 |
Exhaust Filter | H14 HEPA, 0,3 μm particle %99.995< filtration efficiency | |
Main Filter | H14 HEPA, 0,3 μm particle %99.995< filtration efficiency | |
Work Chamber Cleanroom Class | EN ISO 14644-3 | < ISO 5 |
US FED 209E | < Class 100 | |
Work Bench | Standard | 304 Stainless Steel |
Optional | 316L Stainless Steel | |
Noise Level | Normal Mode | ≤ 56 dB(A) |
Stand-by Mode | ≤ 50 dB(A) | |
Front Glass | ≤ 10 mm Laminated | |
Front Glass Glove Port Quantity | 4 | |
Illumination | 850 – 1250 Lux | |
Energy consumption 230 VAC - 50 Hz | Fan Filter Unit | 200 W |
Exhaust Motor | 90 W | |
Illumination | 36 W | |
UV Lamp | 40 W | |
Decontamination Unit | 2800 W | |
Electrical Socket Power / Current | 2x1150 W / 2x5 A | |
Total Power / Current | 5466 W/23.7 A | |
Energy Consumption in Normal Operating Mode (Lighting and Fan Motor) | 326 W | |
Cabinet Weight (kg) | 400 | |
Packaged Weight (kg) | 475 | |
Pack/Palette Dimensions (Included Pass-Box and Support Stand ) | 2405x1150x2000 |