Cytotoxic Safety Cabinet
Metisafe® Cytotoxic work cabinets works under principles with full exhaust of Pre/HEPA filtered fresh air to outer athmosphere without recirculation. Exhausting through HEPA filter protects also environment. Those airflow principles provide safe work with cytotoxic / antineoplastic materials beside of biologicals. Non- recirculated airflow cytotoxic cabinet also prevents possible dangers against volatile vaporizing chemical substances. The room air is pulled in through the perforations located at front of the work plate and directed to blower plenum (inflow air). This airflow direction prevents reaching dirty air to materials over the working plate. Pre/HEPA filtered air supplied through top of the cabinet provides the work area protecting the work zone under ISO-5 class laminar airflow. Total air in the working cabinet fully exhausted without recirculation through exhaust blower fan to athmosphere by keeping working area under negative air pressure. In addition to primary air barrier within the working zone negative plenum design of the Metisafe® cabinet prevents particle escape to surroundings and zero leakage is ensured.
Main Body and Design Features
- Sealed main body covered with Antibacterial epoxy powder paint coated steel
- <7° angle back sloped front panel for work comfort and increased working chamber vision
- 26° Openable front panel for easy working chamber cleaning
- Easy reachable electronic control & electrical part located inside at front panel isolated from work chamber
- UV resistant glass
Microprocessor Control Unit
- Password protected user control,
- Power on/off,
- Standby/Normal mode operation button,
- Motorized front window level position setting and one key sash window preset position,
- UV and FL on-off button,
- UV activation and automatic switch to standby mode only after fully closed front window,
- User interface password change,
- Password protected technical service maintanance / calibration,
- Date and time information
Audio-Visual Warning System
- Unproper front window position
- Unproper air velocity
- Filter and UV lamp replacement
- Technical service maintenance warnings
- Alarm silence button
Quality And Certifications
- CE
- Biological Safety Cabinets EN12469
- ISO9001-2015 Quality Management Certificate
- ISO14644 Cleanroom Class Compatibility
- EN61010 Electric and Electronic Safety Compatibility
- Accredited Testing Institution Validation Guarantee
Optionals
- MOP.ADI Lux Adjustable Illumination
- MOP. FWS Front Window Motion Safety System
- MOP.HTS Humidity/Temperature Sensor
- MOP.RES Remote Access
- MOP.CFL.EU4 Active Carbon Impregnated Pre-filter
- MOP.FLT.ULP ULPA Filter
- MOP.PMS Particle Measure System
- MOP.BWT Precision Balance with Vibration Damping Weighing Table
- MOP.MIE Microscope Entegration
- MOP.WCS Working Cabinet Suspension Apparatus
- MOP.MDS Manual Decontamination System
875x520x746
995x735x1600
995x735x745
≥ 0.40 m/s
≥ 0.40 m/s
0.25 - 0.50 m/s
H14 HEPA, 0,3 μm 0,3 µm paticle %99.995 < filtration efficiency
EU4
304 Paslanmaz Çelik
316 Paslanmaz Çelik
≤ 65 dB(A)
750 Lux
1250x610x675
1370x843x1700
1370x843x745
840 mm
≥ 0.40 m/s
0.25 - 0.50 m/s
H14 HEPA, 0,3 µm paticle %99.995 < filtration efficiency
EU4
304 Stainless Steel
316 Stainless Steel
≤ 65 dB(A)
750 Lux
1480x520x746
1600x735x1590
1600x735x745
840 mm
≥ 0.40 m/s
0.25 - 0.50 m/s
H14 HEPA, 0,3 µm paticle %99.995 < filtration efficiency
EU4
304 Stainless Steel
316 Stainless Steel
≤ 65 dB(A)
750 Lux
1785x520x746
1905x735x1590
1905x735x745
840 mm
≥ 0.40 m/s
0.25 - 0.50 m/s
H14 HEPA, 0,3 µm paticle %99.995 < filtration efficiency
EU4
304 Stainless Steel
316 Stainless Steel
≤ 65 dB(A)
750 Lux